The DL410 is a larger capacity spray dryer that also does not require the liquid sample or solution to undergo any pre or post-processes such as filtration, separation, or pulverization. The use of organic solvents is fully supported with the attachment of our GAS410 organic solvent recovery unit. Small, expensive and/or heat sensitive samples can be dried quickly and efficiently with this easy to operate system. Processes samples as small as 0.5 g of solid matter Safe for heat-sensitive samples, such as food or medical products No risk of contamination Digital display of inlet/outlet temperature and drying air volume Detachable drying chamber, cyclone and product vessel Fast and easy clean up Universal power supply and multilingual touch screen controller
The hot air inlet and drying chamber cover automatically move up and down, and since the cyclone and product vessel can easily be removed, cleaning and maintenance after your experiment is easy Control functions are conveniently arranged on the control panel for various conditions The temperature recorder, air flow meter, pressure gauge and other measurements allow easy control of experiment conditions