Microcomputer PID control system enables high-precision temperature control and is equipped with programmed operation.
It is equipped with a self-diagnosis function for trouble with temperature regulator, temperature sensor, SSR, heater, level sensor, etc. In addition, it is equipped with an overheat prevention device and an earth-leakage breaker for safe continuous operation.
Equipped with an external circulation pump that supplies constant hot water.
Temperature display and setting are easy to read digital. Specifications
Specification
Operating Temperature Range: -10°C to +100°C
Temperature adjustment accuracy: ±0.05°C (at 20°C)
Required Tank Fluid Amount: 4.5 L to 5.5 L * Medium Fluid: Water and Ethylene Glycol Brine (Nybran Z-1 Recommended)
Tank Opening: 110 (W) x 110 (D) x 130 (H)
External Dimensions (mm): 200 (W) x 395 (D) x 570 (H)
External Circulation Hose Port: Outer Diameter 8 mm
Circulation pump: Maximum flow rate approx. 4 L/min, maximum lift distance 1.4 m
Temperature Controller: PID Control by Microcomputer
Operation mode: fixed value operation, program operation
Display and settings: 4 digit display (temperature display resolution: 0.1°C) Digital settings with key switches
Heater: Stainless sheath heater
Refrigerator: Air-Cooled Sealed 100 W (Off at 38°C±3°C)