Measurement object: Substances that exhibit infrared absorption at a wavelength of around 3.4~3.5 μm extracted from sample water in an oil extraction solvent
Display format: 3.5 inch 320 x 240 dot backlit color graphic LCD
Calibration Method: Zero, Span Calibration: Automatic or Manual Calibration after Liquid Injection
Volume of sample water collected: 2:1 (sample: oil extraction solvent)
Extraction solvent: H-519 * Do not use oil extraction solvent other than H-519.
Extracting solvent: 8 mL (or 10 mL)
Extraction method: by the extraction function included in the analysis section
Ambient Temperature: 0~40°C (non-condensing)
Power: 100~240 VAC ±10%%%% 50/60Hz
Power Consumption: Approx. 60VA at 100 VAC, Approx. 90VA at 240 VAC
External Dimensions (mm): 200 x 313 x 342
Weight: approx. 7 kg
External output: Output to USB thumb drive
Measurement flow: After liquid injection, automatic measurement (automatic switching sequence) and manual measurement
Others: 300 data memory function, self-discrimination function, measurement stability value display function, clock function, agitator backlight, unit conversion function, ASTM D7066-4 compliant, density correction function