With the program control specification of TRL-108H, it can be used for a wider range of applications.
Simple Configuration Operations
▼ Programs of up to 20 steps can be run easily by simply setting items displayed on the message board. (No troublesome parameter settings are required.)
Selectable operation and program operation are possible.
▼ It is possible to choose the operation method, so of course, conventional fixed price operation is also possible.
Specification
External Dimensions: W265 x D375 x H455mm
Tank Inner Dimensions: W180 x D110 x H180mm
Content: 3L
Temperature Range: -10°C to +80°C
Temperature stability: ±0.05°C (in tank)
Temperature control: digital program control PID control system
▼ program operation constant value operation switching type
▼ Number of stored patterns: 1 pattern
▼ Number of stored steps Max. 20 steps/1 pattern
▼ Setting time range 99 hours 59 minutes/1 step
▼ Selection of cooling operation (refrigerator ON-OFF) for each step
Heater: Sheath Wire Type 500 W
Refrigerator: Air-Cooled 130 W HFC-134a
Cooling Capacity: Approx. 180 W (155kcal/h Water Temperature +10°C 50 Hz)
Stirrer: by jet of discharge liquid
Circulation Pump (Also Used for Stirring): 20 W Magnet Pump
▼ Maximum lift 2 m
▼ Maximum flow rate 9 L/min
▼ Maximum discharge pressure 0.02Mpa
Circulation IN-OUT: Outer Diameter φ12 Hose Port (Rp 3/8)
Safety devices: earth leakage breaker, self-diagnostic function (HC: Excess Rise Prevention, LC: Low Temperature Prevention, Water Level Reduction, Refrigerator Protection Circuit, Temperature Sensor Abnormality, Refrigerator Pressure Abnormality, Missetting Prevention)
*When the temperature in the tank is 40°C or higher, the refrigerator is forcibly stopped by the action of the refrigerator protection circuit. For this reason, refrigeration cannot be used in the temperature range of 80°C to 40°C.