Calibration gas can be continuously generated for a long period of time
The concentration adjustment range of calibration gas is wide and easy to generate
High reliability because the concentration is determined by measuring the basic physical quantity of weight loss and dilution gas
Many kinds of trace concentration calibration gas can be generated
The dilution gas flow rate is automatically calculated by inputting the desired gas concentration on the touch screen
Control the flow rate change by installing a mass flow controller (gas flow controller)
Specification
Generation target gas permeation tube method: Gas sold as permeation tube
Diffusion tube method: A stable, high-purity material having a vapour pressure in the range 5-400 mm Hg at a temperature in the set range of PD-1C as a characteristic of the liquid.
Dilution gas type Nitrogen or air
Dilution gas pressure: 0.15-0.6 MPa
Flow setting range: 0.2-10 L/min
Flow setting accuracy
▼ 0.2-0.4L/min:±13.2%%%%s.p.
▼ 0.4-1L/min:±7.0%%%%s.p.
▼ 1-10L/min±3.7%%%%s.p.
Temperature setting range: (room temperature +5°C)-50°C (room temperature is 15-30°C)
Temperature setting accuracy: ±0.15°C
Tube holder capacity
▼ Permeation Tube: Standard Size 10, H Size 5 (Permeation tube having H in catalog No. and P-17-5)
▼ Diffusion Tube: 1 for Single Component, 4 for Multiple Components
Power Supply/Maximum Power Consumption: 100-240 V AC 50/60Hz 160 W
Dimensions: 250 (W) x 340 (H) x 315 (D) mm
Weight: approx. 14 kg (without water)
Accessories: 2 Calibration Gas Tubes (Teflon Tube Outer Diameter φ6 x Inner Diameter φ4 mm2 m), 1 Dilution Gas Tube (With Dust-Proof Filter) (Outer Diameter φ6 x Inner Diameter φ4 mm2 m), 1 Power Cord (2 m), 1 Power Conversion Adapter, 1 Rubber Plug, 1 Tube Holding Basket
Compliance Planning Constant Voltage Directive: EN61010-1 EMC Directive: EN61326-1 (For use in industrial locations)