Temperature Display Range/Resolution: 0 ~ 50 °C/0.1 °C
Calibration points: Maximum 3 points
Resolution
(1)0 ~ 199 μS/cm: 1 μS/cm
(2)200 ~ 1999 μS/cm: 1 μS/cm
(3)2.00 ~ 19.99 mS/cm: 0.01 mS/cm
(4)20.0 ~ 199.9 mS/cm: 0.1 mS/cm
Accuracy: ± 2% F.S. for range (1) - (3)/± 5% F.S. for range (4)
Main functions: Standard Liquid Automatic Recognition/Automatic Calibration (1 ~ 3 items selectable) * High Electrical Conductivity Standard Liquid (514 -23)/Automatic Temperature Conversion (25 °C Temperature Conversion Factor 2% Fixed)/Temperature Sensor Adjustment/Unit Change (S/cm ⇔ S/m)/Automatic Range Switching/Auto Hold/Auto Stability/Auto Power Off (30 minutes)/Battery Level Indication/Sensor Exchangeable/Waterproof/Dustproof Structure (IP 67 Equivalent)/TDS Conversion Function (Factor (0.4 ~ 1.0), 442, NaCl, EN 27888)