TOKYO RIKAKIKAI CO., LTD. 具有熱量計的工藝反應器 DDS-2000A DDS-2000A
Process Reactor with Calorimetry DDS-2000A
詢價商品請諮詢取得報價及交期
商品包含多種型號及規格,請點選「立即選購規格」挑選
付款方式
月結報帳 / 信用卡 / 銀行轉帳
運送方式
宅配快速到貨,滿 NT$ 999 元即享免運!
選購型號
商品介紹
Feature
For scale-up safety assessment and process optimization experiments • With software suitable for process optimization and scale-up, it is possible to control individual reaction parameters with high precision, perform process optimization, and perform scale-up prediction and scale-up safety assessment.
The original split aluminum block thermostatic bath has the same heat transfer performance as the oil jacket type. Has excellent operability with heating and cooling functions.
Specifications
Impeller stirring
Reaction vessel/number: Separable flasco/1 bottle
Synthetic scale: 250 mL/500 mL/1 L/2 L (choice)
Temperature adjustment range: - 20 ~ 200 C (aluminum block temperature) * A cooling water circulator is required.
Maximum heat removal: 0.50 W/g (with 1 L of water in the glass reactor)
Setting and viewing: Using a laptop
Temperature Control: Heating and Cooling P.I.D, Cascade, Reactor, Jacket, Insulation, Isothermal, Gradient, 8-Step Program
Dropping control: Volume drop or weight drop setting, drop time, drop speed, upper and lower temperature limit, drop temperature step, 1 ~ 2 rate
Stirring control: Speed feedback control, 8-step program (gradient speed + constant speed), impeller (S-type impeller, three-back impeller, half-moon impeller, four-pitch buddle impeller), rotation direction CCW or CW
pH control (optional): 0 ~ 14 pH, acid or base (1 or 2 liquid control)
Calorimetry: Heat flow, specific heat and reaction heat calculation, calibration heater, with CalGraph for recalculation and analysis
Setting and recording: Setting and recording temperature for each experiment, automatic recording of dripping and other parameters
Trend graph export: Real-time pen, with function for retrieving data displayed and stored in historical pen to Excel
Safety functions: Automatic activation during emergency cooling function/reactor control, end stop setting, with emergency program setting, earth leakage/overcurrent breaker, aluminum block independent overrise detector, motor overload protection, software self-diagnosis function (Upper temperature, temperature rise speed, motor overload), communication error
Reaction observation: Aluminum block with front observation window, white LED backlight
Operating ambient temperature range: 5 ~ 35 degrees C
Wetted part materials: Glass, Teflon (R), Hastelloy (R) C
External dimensions: reactor unit/419 x 470 x 999 mm, control box/309 x 362 x 167, AC relay unit/200 x 150 x 89 mm
Power input/voltage: 14.5 A, 2.9 kVA/200 V AC, single-phase 50/60 Hz
*Power plug not included.
*A laptop is included.
Performance is based on room temperature of 20 degrees Celsius, rated power supply voltage of 50 Hz and cooling water circulator CA -1116 A (55% Naive Line (R), Circulation Temperature -20 deg.).