With a temperature setting range of -20~30 degrees Celsius, it can be used in a variety of applications. With a compact footprint of 205 W x 405Dmm, it takes up no space when installed in the draft chamber, above, below or beside the laboratory table.
Has an alarm signal output jack. Upon receiving the alarm signal, the circulating device can be stopped or a commercially available patrol light can be turned on.
This product complies with the RoHS Directive.
Specifications
Cooling/Circulation System: Air Cooled/Sealed System Directional Circulation
Temperature setting range: - 20 ~ 30 C (without heater)
Temperature control accuracy: +/- 2 degrees C (0 ~ 20 degrees C setting), +/- 1 degrees C (- 20 ~ 0 degrees C setting)
Cooling capacity (at liquid temperature): 10 degrees C/450 W (387 kcal/h), -10 degrees C/310 W (266 kcal/h)
External circulation capacity (50/60 Hz): maximum flow rate 9/10 L/min, maximum head 4.2/5.6 m
Temperature control: Refrigerator ON-OFF control
Temperature setting/display: Sheet key input/digital display, minimum digit 1 degree C
Safety functions: Leakage/overcurrent breaker, refrigerator overload relay holding circuit, temperature controller self-diagnostic function (Refrigerator malfunction, power failure alarm, sensor malfunction), refrigerator protection timer, circulation pump impedance protection